PID correction factors for 4 instrument families · Measurement uncertainty per ISO GUM with EN 689 decision rules
Correction Factor (CF) / Response Factor (RF) — PIDs are calibrated to isobutylene (CF = 1.0). To obtain the true concentration of another compound:
True concentration = PID reading × CF.
CF > 1 means the instrument underreads; CF < 1 means it overreads.
NR = no response (compound ionisation energy exceeds lamp energy).
Search for a compound above to view correction factors across manufacturers.
Enter the compound, lamp energy, and PID reading to calculate the true concentration. Select the manufacturer whose CF values you want to apply.
Compound
CAS
IE (eV)
RAE 10.6
RAE 9.8
RAE 11.7
MX6 10.6
IGD 10.6
Drg HC
Drg LC
How PIDs work
A photoionization detector (PID) ionises gas molecules using UV light and measures the resulting ion current. PIDs are calibrated to isobutylene (2-methylpropene, CAS 115-11-7), assigned a CF of 1.00. Because different compounds ionise with different efficiency under the same lamp, a correction factor is required to convert the displayed reading to the true concentration.
Correction Factor (CF) = Response Factor (RF) — The terms are used interchangeably across manufacturers. RAE Systems uses CF; MSA, Industrial Scientific, and Dräger use RF. The calculation is identical: True concentration = PID reading × CF.
NR (No Response) — The compound's ionisation energy (IE) exceeds the lamp photon energy. The compound cannot be detected with that lamp.
Lamp energies: The most common PID lamp is 10.6 eV. High-sensitivity lamps at 11.7 eV can detect compounds with higher IE (e.g. some halogenated solvents), but degrade faster and are humidity-sensitive. Low-energy 9.8 eV lamps reduce sensitivity to common interferences.
Dräger sensor variants: The PID HC (high concentration) is suited to ppm-range work; the PID LC ppb version has a smaller measurement cell for sub-ppm detection. RF values differ between the two.
Manufacturer
Instrument
Lamp (eV)
Source
Compounds
RAE Systems
MiniRAE, ppbRAE, MultiRAE
9.8 / 10.6 / 11.7
TN-106
262
Industrial Scientific
MX6 iBrid
10.6
PID Response Factors table
115
IGD / GenSenova
SL-031
10.6
CF table SL-031-2
141
Dräger
X-am 8000
10.6
Substance list (PID HC + LC ppb)
90
MSA
ALTAIR 5X
10.6
Compound library (RF device-stored)
118 (no published RF)
CF values are sourced directly from manufacturer technical documentation. Values are for informational purposes only; always verify against the current manufacturer datasheet for your specific instrument and lamp. Humidity, temperature, lamp age, and calibration gas concentration all affect accuracy. This tool does not replace site-specific calibration procedures.
ISO GUM approach — occupational exposure measurement
Enter the uncertainty components for your measurement. Each component is expressed as a relative standard uncertainty (%). The combined standard uncertainty uc is calculated by root-sum-of-squares (RSS). The expanded uncertainty U (k=2, ~95% confidence) is used for OEL comparison. Inputs follow EN 482:2021 / EN 13936:2014 framework.
Measurement result
Uncertainty components (relative standard uncertainty, %)
Enter components above to build the uncertainty budget.
Combined and expanded uncertainty
Combined standard uncertainty uc
uc = √(us² + ua² + uv² + uw² + ux²)
—
Expanded uncertainty U (k=2)
U = k × uc
—
Measured value
—
Measurement interval [C − U, C + U]
—
OEL
—
Enter measured value and OEL to assess compliance with uncertainty.
Combined standard uncertainty: uc = √Σui² (relative). Expanded uncertainty U = k × uc. Compliance interpretation per EN 13936:2014 and EN 689:2018: if C + U ≤ OEL, the measurement is fully compliant; if C − U > OEL, the OEL is exceeded with high confidence; if the interval straddles the OEL, further sampling is required. Standard: ISO/IEC Guide 98-3:2008 (GUM); EN 482:2021; EN 689:2018; EN 13936:2014.
Typical uncertainty values for occupational sampling methods
Values below represent typical combined standard uncertainties for published NIOSH/OSHA methods under controlled conditions. Actual values should be taken from method validation data or laboratory accreditation certificates. Workplace variability is excluded from these figures.
Method reference uncertainty values
Method / analyte
Sampling us
Analytical ua
Volume uv
Combined uc (without variability)
NIOSH 1500 — Hydrocarbons (GC-FID)
5%
3.5%
3%
~7%
NIOSH 2010 — Ketones (GC-FID)
6%
4%
3%
~8%
NIOSH 2531 — Formaldehyde (HPLC)
8%
5%
3%
~10%
NIOSH 5000 — Inorganic dust, gravimetric
7%
2%
4%
~8%
NIOSH 7400 — Asbestos fibres (PCM)
10%
20%
3%
~22%
NIOSH 9000 — Endotoxin (LAL assay)
15%
25%
5%
~30%
OSHA 7 — Mercury vapour (AAS)
5%
5%
3%
~8%
IOM inhalable dust — gravimetric
9%
2%
4%
~10%
GSP respirable cyclone — gravimetric
8%
2%
4%
~9%
Direct-reading instruments (calibrated)
10–25%
—
—
10–25%
EN 13936 / EN 689 compliance decision rules
Condition
Decision
Required action
C + U ≤ OEL
Compliant
No immediate action; maintain monitoring programme
C − U ≤ OEL < C + U
Inconclusive — further sampling required
Collect additional samples; consider 8-sample strategy per EN 689:2018 Annex G
C − U > OEL
Non-compliant
Implement controls; re-measure after controls to verify compliance
References: ISO/IEC Guide 98-3:2008 (GUM); EN 482:2021; EN 689:2018; EN 13936:2014. NIOSH manual: www.cdc.gov/niosh/nmam/